Vue de la page 0
PanOptic Ophthalmoscope
118 Series
Directions
for use
Vue de la page 0
1 2 3 4 5 6 ... 27 28

Résumé du contenu

Page 1 - PanOptic™ Ophthalmoscope

PanOptic™ Ophthalmoscope118 SeriesDirectionsfor use

Page 2

6Attaching Corneal Viewing Lens to PanOptic (Model 11820 only)Insert Corneal Viewing Lens into patient’s side of the instrument. Pushand twist in one

Page 3 - PanOptic

7Conducting an Examination with the PanOpticOphthalmoscope1. To examine the right eye, stand or sit to the patient’s right side.2. Remove Spectacles (

Page 4

85. Turn the scope on (using anyWelch Allyn 3.5V power source).Adjust light intensity rheostat todesired position.You are now ready to begin the exami

Page 5

910. Slowly follow the red reflex towardthe patient and into the pupil. Getas close as possible for theoptimum view (the eyecup shouldcontact the pati

Page 6 - Warnings and Cautions

1012. Examine the optic disc for clarity ofoutline, color, elevation andcondition of the vessels. Followeach vessel as far to the peripheryas you can.

Page 7

1114. To examine the left eye, repeat theprocedure outlined above.However, unlike traditionalophthalmoscopy, a unique featureof PanOptic is that the e

Page 8

12Apertures and FiltersThere is a wide range of practical apertures and filters to select from:small spot, large spot, micro spot, slit aperture, red-

Page 9 - Set-up Instructions

13used to help distinguish veins from arteries; veins stay relativelyblue, but oxygenated arterial blood makes arteries appear blacker.This greater co

Page 10

14The EyeIn addition to examination of the fundus, the PanOpticOphthalmoscope is a useful diagnostic aid in studying other ocularstructures. The light

Page 11 - Ophthalmoscope

15Lamp Replacement InstructionsWARNING: Lamps may be hot. Lamp should be allowed tocool before removal. Use only Welch Allyn model 03800-Uor 03800-LED

Page 12

iiPatent informationFor patent information, please visit www.welchallyn.com/patents.©2013 Welch Allyn, Inc. To support the intended use of the product

Page 13

16RepairsRepairs should be performed at the factory.Cleaning and DisinfectionLenses: PanOptic is a precision optical instrument. Do not use anysolvent

Page 14

17TroubleshootingTrouble Area Possible CauseCorrectiveActionNo Light outputAperture dial is in-between positionsRotate the aperture dial.Lamp burned o

Page 15

18Dim Light OutputLamp pin not engaged inslot.Insert the lamp with pinfirmly seated in slot.Film on Optics. Clean the lenses with aclean/soft optical

Page 16 - Apertures and Filters

19Accessory and Replacement Parts ListItemNo.Description Product #1 Corneal Viewing Lens(model no. 11820 only)118752 Patient Eyecups (Package of five

Page 17

20Welch Allyn Technical Support:http://www.welchallyn.com/about/company/locations.htmPanOptic Ophthalmoscope WarrantyWelch Allyn, Inc. warrants the No

Page 18

21SpecificationsModel Number11810 PanOptic Ophthalmoscope without Cobalt Blue Filter11820 PanOptic Ophthalmoscope with Cobalt Blue Filter and Add-on C

Page 21 - Troubleshooting

24Welch Allyn, Inc.4341 State Street RoadSkaneateles, NY 13153-0220Tel: (800) 535-6663www.welchallyn.comRegulatory Affairs RepresentativeWelch Allyn L

Page 22 - Corrective

Table of ContentsSymbols ...1Warnings and Cautions ...2The PanOptic™Ophthalmoscope ...

Page 23 - Description Product #

ivPanOptic Ophthalmoscope Warranty ...20Specifications...21

Page 24

1SymbolsAttention. Read Operating Manual for Cautions and Instructions for Use.Type BF applied partMeets essential requirements of European Medical De

Page 25 - Specifications

2Warnings and CautionsAuthorized representative in the European CommunityGlobal Trade Item NumberWARNING: The light emitted from the ophthalmoscope is

Page 26

3WARNING: To minimize lamp housing temperatures, on-time should not exceed 2 minutes with off-time not less than10 minutes.WARNING: Use only PanOptic

Page 27

4The PanOptic™OphthalmoscopeTransparency of the cornea, lens and vitreous humor permits thepractitioner to directly view arteries, veins, and the opti

Page 28

5Set-up InstructionsAttaching the Pan OpticOphthalmoscope head to a WelchAllyn Power SourceAlign cutouts in the PanOpticOphthalmoscope base with lugs