CP 150 12-lead restingelectrocardiographQuick reference guide
Item Button1 Leads button2 ECG preview button3 Gain button (size)4 Speed button5 Filters button7. If an Artifact message appears, minimize the artifac
11. After the test prints, select the desired option: Reprint, Save, Retest, or Rhythm. Ifthe Auto Save setting is turned off, touch Save to save the
10 Performing an Auto ECG test CP 150 12-lead resting electrocardiograph
TroubleshootingLead-quality problems“Artifact” message on the screenArtifact is signal distortion that makes it difficult to accurately discern the wa
Muscle tremorCauses• The patient is uncomfortable, tense, nervous.• The patient is cold and shivering.• The exam bed is too narrow or short to comfort
Lead alert or square waveA dot might be flashing on the lead-status screen. Or one or more leads might appear asa square wave.Causes• Electrode contac
Material No.721331
© 2013 Welch Allyn, Inc. To support the intended use of the product described in this publication, the purchaser ofthe product is permitted to copy th
Powering the electrocardiographThe electrocardiograph runs on AC or battery power. Plug the electrocardiograph into ACpower as often as possible so th
WARNING Do not allow the conductive parts of the patient cable,electrodes, or associated connections of defibrillation-proof applied parts,including
Attaching the leads to the patientProper lead attachment is important for a successful ECG. The most common ECGproblems are caused by poor electrode c
Electrode examples, left to right: arm clamp (reusable), Welsh cup (reusable), tabelectrode (disposable), monitoring electrode (disposable).• For reus
ECG home screenECG home screenThe ECG home screen includes the following areas:Item Area1 Device status2 Content3 NavigationDevice status areaThe Devi
Content areaThe Content area includes 3 test selection buttons and a preview selection button:• Auto ECG• Rhythm ECG• Stat ECG• Electrode PlacementThe
Performing an Auto ECG testCAUTION Patient data is not saved until the ECG test is completed.Note The ECG configuration settings can be changed in th