TESTDANGER0.10.20.30.40Cleaning (continued)Perform Leak Test ProcedureThe leakage test allows for a simple two stage test of the water tight integrity
Cleaning (continued)3. Pressurize the scope by pumping the hand bulb until the indicator on the gauge is in the green zone. DO NOT pressurize into the
Cleaning (continued)NOTE: Never connect or disconnect the leakage tester under water.This will allow water into the instrument and leakage tester.4. I
Disinfection (continued)2. The instrument should remain in contact with the disinfecting solution for the time period recommended by the manufacturer
Disinfection (continued)Disinfecting SolutionsSpecific reference to brand names is not an endorsement of their efficacyas a disinfecting solution. Tes
Sterilization And Aeration (continued) 2. Following ETO Gas Sterilization, aeration time of 72 hours at room temperature is required.3. Aeration Chamb
4341 State Street RoadSkaneateles Falls, NY 13153315-685-4560800-535-6663Printed in U.S.A. Part No. 650033 Rev. C
Thank you for purchasing the Welch Allyn RL-150™ Rhino-laryngoscope. The operating and maintenance instructionsfound in this manual should be followed
SpecificationsAngle of View ...75°Depth of Field...3-50 mmDiopter...+2 to -8 diopterTip De
NomenclatureRL-150 Rhinolaryngoscope2LX-150 Light SourceDiopterAdjustmentRingAngulationLeverUmbilicalCordPowerSwitchAirSwitchIlluminationControlEndosc
Prior to Initial UseBefore set up and inspection of the equipment, check all componentsreceived against the list of components (see Components section
System Set Up (continued)Light Source (See Light Source Manual for complete instructions.)1. Plug power cord into the receptacle on the back of the ca
System Inspection (continued)Inspection of Light Guides and Optics1. With the light source lamp on, hold distal tipapproximately 40 mm from any printe
OperationProcedureThe methods and techniques of rhinopharynlaryngoscopy are well-definedand documented. Endoscopy training seminars and programs are a
CleaningEndoscopic instruments should be cleaned immediately after each use.Endoscopes are delicate and will degrade if not cleaned promptly, due toth