Module OEM Implementation Manual
Page 24 Confidential Welch Allyn OEM Technologies
Troubleshooting
Make sure that the room is well ventilated. CO
2
readings may be elevated in a closed
room.
Confirm that the calibration gas is of known concentration. If the CO
2
readings are
consistently elevated or depressed, the calibration gas may be suspect. Try another
gas source.
Make sure that the CO
2
module was properly set up and operating for the required
minimum period before testing.
Check for air leaks in the pneumatic tubing and connections. Repair or replace as
needed.
Confirm that the sample gas flow rate is greater than the CO
2
module’s pump flow
rate.
Confirm that silicon tubing is not used in the procedure (silicon absorbs CO
2
which
affects the gas concentration stability and accuracy of the procedure).