Welch-allyn LC101 CO2 Module - User Manual Manuel d'utilisateur Page 71

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 98
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 70
LC101 CO
2
Module OEM Implementation Manual
Appendix C - Error Messages & Recovery
Welch Allyn OEM Technologies Confidential Page 71
57, CO
2
Sensor Not Found
CO
2
Bench, or C-Cap, not present on Main Board.
60, CO
2
Sensor EEPROM Error/Revision #
Revision # read from the CO
2
sensor EEPROM is not in the 1.xx format.
65, CO
2
Sensor EEPROM Error/Read/Write
A read or write error to the CO
2
sensor EEPROM has occurred.
66, CO
2
Sensor EEPROM Error/CCITT/CRC Code
CCITT/CRC code read from the CO
2
sensor EEPROM is invalid.
70, CO
2
Sensor Temperature Too High
The CO
2
sensor’s temperature is greater than 60
o
C for more than 30 seconds.
Check for the presence of an external excessive heat source that is in close proximity
to the Module. Remove the source or relocate the Module.
71, CO
2
Sensor Temperature Too Low
The CO
2
sensor’s temperature is less than 5
o
C for more than 30 seconds.
Check for excessive cold ambient conditions or the presence of an external cold
source that is in close proximity to the Module. Remove the source or relocate the
Module.
Vue de la page 70
1 2 ... 66 67 68 69 70 71 72 73 74 75 76 ... 97 98

Commentaires sur ces manuels