Propaq® EncoreVital Signs MonitorReference GuideModel 202EL, 204EL, 206ELSoftware version 2.4X
Safety SummaryWelch Allyn 10 Propaq Encore Reference Guide Warning Make frequent electrical and visual checks on cables and electrode wires. Avoid
Setting Alarm LimitsWelch Allyn 100 Propaq Encore Reference GuideCustomizing Propaq Encore Alarm LimitsAlthough adjusting the Propaq's alarm limi
Equipment AlertsPropaq Encore Reference Guide 101 Welch AllynEquipment AlertsDefinitions and IndicationsThe Propaq Encore can alert you to changing equ
Equipment AlertsWelch Allyn 102 Propaq Encore Reference GuideIf the equipment condition also caused a patient alarm, you will need to first suspend the
Propaq Encore Reference Guide 103 Welch AllynTrendsEvery two minutes, the Propaq Encore collects the monitored vital sign numerics and stores them in
How Trends are AccumulatedWelch Allyn 104 Propaq Encore Reference GuideHow Trends are AccumulatedFor each of the four continuous trends, numerics are
Displaying TrendsPropaq Encore Reference Guide 105 Welch AllynSelecting a Trend (NXT TRND)Select the trend you want displayed by pressing the NXT TRND
Displaying TrendsWelch Allyn 106 Propaq Encore Reference Guide
Propaq Encore Reference Guide 107 Welch AllynPrintingPrinting Patient DataPrinting WaveformsIf you have an expansion module set up with your monitor,
Printing Patient DataWelch Allyn 108 Propaq Encore Reference GuideCO2 and RESP WaveformsCO2 waveforms are printed on a grid with major divisions (vert
Printing Patient DataPropaq Encore Reference Guide 109 Welch AllynPrinting NIBP MeasurementsYou can print the results of an NIBP measurement each time
Safety SummaryPropaq Encore Reference Guide 11 Welch Allyn The Propaq Encore should be serviced only by a Welch Allyn service technician while under
Printing Patient DataWelch Allyn 110 Propaq Encore Reference GuidePrinting the Apnea TicketThe Apnea Ticket documents the length of each apnea episode
Printing Patient DataPropaq Encore Reference Guide 111 Welch AllynPrinting When a Patient Alarm OccursTo set up the printer to print on a patient alar
Printing Patient DataWelch Allyn 112 Propaq Encore Reference GuideOxyCRGThe OxyCRG is a graphical printout of two minutes of continuous HR/PR and SpO2
Printing TrendsPropaq Encore Reference Guide 113 Welch AllynPrinting TrendsPrinted trends are useful for reviewing the patient's vital signs over
Printing TrendsWelch Allyn 114 Propaq Encore Reference GuideAutomatic Trend Prints To automatically print trends at 4-hour intervals, activate AUTO TR
Propaq Encore Reference Guide 115 Welch AllynAcuity Central Monitoring SystemIntended UseThe Acuity system is used as a central monitoring system for
Connecting to the Acuity SystemWelch Allyn 116 Propaq Encore Reference Guide2. If the Propaq Encore is not already connected to the patient, attach le
Press NET OFF to Disconnect from AcuityPropaq Encore Reference Guide 117 Welch AllynKey-press Route to Acuity MenuPress NET OFF to Disconnect from Acu
Printing at AcuityWelch Allyn 118 Propaq Encore Reference Guide2. Within 15 seconds, disconnect the Acuity network cable from either the Propaq Encore
Propaq Encore Reference Guide 119 Welch AllynPower SourcesFor in-hospital operation and recharging from ac mains, an ac power adapter plugs into the m
Safety SummaryWelch Allyn 12 Propaq Encore Reference Guide Off (Standby) Patient connections are Type CF, isolated for direct cardiac application
Power Adapter Intended UseWelch Allyn 120 Propaq Encore Reference GuideVerifying Proper Power Adapter ConfigurationPrior to using the power adapter, ch
Power Adapter Intended UsePropaq Encore Reference Guide 121 Welch AllynProcedure1. Unplug the power adapter’s removable cord from the ac mains outlet
Power Adapter Intended UseWelch Allyn 122 Propaq Encore Reference GuideCaution Make sure the voltage selector indicates the proper ac input voltage.
Battery CarePropaq Encore Reference Guide 123 Welch AllynBattery CareRecharging TimeThe battery charges to full capacity within eight hours (if the mo
Battery CareWelch Allyn 124 Propaq Encore Reference GuideMonitor Functions Based on Battery VoltageAs battery voltage drops during extended monitor ba
Propaq Encore Reference Guide 125 Welch AllynCare and MaintenanceInspect and Clean the Monitor and AccessoriesBefore cleaning, thoroughly inspect the
MaintenanceWelch Allyn 126 Propaq Encore Reference GuideMaintenanceService Interval RecommendationsAt the intervals recommended below, verify the Prop
Monitor CarePropaq Encore Reference Guide 127 Welch AllynProduct RecyclingYou can return a Propaq Encore to Welch Allyn for recycling when the monitor
Printer MaintenanceWelch Allyn 128 Propaq Encore Reference GuidePrinter Paper RemovalPrinter MaintenanceLoading PaperPaper is loaded through the botto
Printer MaintenancePropaq Encore Reference Guide 129 Welch Allyn5. Slide the end of the paper into the slot of the printing mechanism until it extends
Propaq Encore DocumentationPropaq Encore Reference Guide 13 Welch Allyn Propaq Encore Documentation The Documentation Set The Propaq Encore documenta
Customer ServicesWelch Allyn 130 Propaq Encore Reference GuideCustomer ServicesOrdering and Customer ServiceFor ordering information, for the location
Propaq Encore Reference Guide 131 Welch AllynDefibrillator SynchronizationThese instructions are not intended to replace existing hospital procedures
Removing the Defib Sync Interface from the LIFEPAK 5Welch Allyn 132 Propaq Encore Reference Guide2. Slide the Propaq Encore Defib Sync interface onto th
Synchronized Cardioversion Using the LIFEPAK 5Propaq Encore Reference Guide 133 Welch AllynSynchronized Cardioversion Using the LIFEPAK 51. Set up the
Synchronized Cardioversion Using the LIFEPAK 6sWelch Allyn 134 Propaq Encore Reference Guide5. Follow hospital procedures and LIFEPAK 5 instructions f
Synchronized Cardioversion Using the LIFEPAK 6sPropaq Encore Reference Guide 135 Welch Allyn2. Verify the integrity of the ECG patient electrodes and
Defib Sync MessageWelch Allyn 136 Propaq Encore Reference Guide7. Follow hospital procedures and LIFEPAK 6s instructions for cardioversion.8. If subseq
Propaq Encore Reference Guide 137 Welch AllynSpecificationsECGThe ECG channel meets all the requirements for Cardiac Monitors Heart Rate Meters and Al
ECGWelch Allyn 138 Propaq Encore Reference GuideQRS Detector Adult or Pediatric Amplitude Range: 0.22 to 5.0 mV (RTI)Neonatal Amplitude Range: 0.1 to
Real-Time ECG Analog/Defib SyncPropaq Encore Reference Guide 139 Welch AllynReal-Time ECG Analog/Defib SyncSpecial cables are required to interface the
Propaq Encore DocumentationWelch Allyn 14 Propaq Encore Reference Guide
Impedance Pneumography (RESP)Welch Allyn 140 Propaq Encore Reference GuideImpedance Pneumography (RESP) Impedance Pneumography (RESP) SpecificationsCha
Invasive PressurePropaq Encore Reference Guide 141 Welch AllynInvasive Pressure Applies only to models 204 and 206.Invasive Pressure SpecificationsChar
NIBPWelch Allyn 142 Propaq Encore Reference GuideNIBPNIBP SpecificationsCharacteristic SpecificationMethod OscillometricControl Automatic and manual mea
TemperaturePropaq Encore Reference Guide 143 Welch AllynTemperatureTemperature SpecificationsMinimum Time between automatic measurements30 seconds (Aut
Pulse Oximetry (SpO2)Welch Allyn 144 Propaq Encore Reference GuidePulse Oximetry (SpO2)Masimo SpO2 SpecificationsCharacteristic SpecificationSaturation
Pulse Oximetry (SpO 2 )Propaq Encore Reference Guide 145 Welch Allyn Nellcor SpO 2 Specifications CharacteristicSpecification a a.When performing SpO2
Capnography (CO2)Welch Allyn 146 Propaq Encore Reference GuideCapnography (CO2)General CO2 SpecificationsCharacteristic SpecificationCO2 DisplayScreen D
Capnography (CO2)Propaq Encore Reference Guide 147 Welch AllynBarometric PressurePressure Compensation AutomaticOperating Range -2,000 to 15,000 ft (-
Capnography (CO2)Welch Allyn 148 Propaq Encore Reference GuideMainstream CO2 SpecificationsCharacteristic SpecificationMainstream CO2 SensorSensor Type
Capnography (CO2)Propaq Encore Reference Guide 149 Welch AllynSidestream CO2 SpecificationsCharacteristic SpecificationSensor Type Sidestream, internalP
Propaq Encore Reference Guide 15 Welch Allyn Getting Started Introducing the Propaq Encore Intended Use Before using the Propaq Encore on a patien
AlarmsWelch Allyn 150 Propaq Encore Reference GuideAlarmsAlarms SpecificationsCharacteristic SpecificationIndicators Flashing red ALARM light indicates
TrendsPropaq Encore Reference Guide 151 Welch AllynTrendsTrends SpecificationsDisplayDisplay SpecificationsCharacteristic Specificationaa.Assumes SpO2 an
Monitor (Environmental)Welch Allyn 152 Propaq Encore Reference GuideMonitor (Environmental)Monitor Environmental SpecificationsCharacteristic Specificat
Monitor (Physical)Propaq Encore Reference Guide 153 Welch AllynMonitor (Physical) Monitor Physical SpecificationsCharacteristic SpecificationProtection
PrinterWelch Allyn 154 Propaq Encore Reference GuidePrinterPrinter SpecificationsCharacteristic SpecificationOperationOperating Modes Continuous, Snapsh
PowerPropaq Encore Reference Guide 155 Welch AllynPowerMonitor Power SpecificationsElectromagnetic Compatibility (EMC) Per IEC 601-1-2, which is a coll
Power AdaptersWelch Allyn 156 Propaq Encore Reference GuidePower AdaptersPower Adapter General SpecificationsCharacteristic SpecificationProtection Clas
Power AdaptersPropaq Encore Reference Guide 157 Welch AllynPower Adapter Physical SpecificationsCharacteristic SpecificationUniversal Power Adapter, Par
Factory Default SettingsWelch Allyn 158 Propaq Encore Reference GuideFactory Default SettingsFactory Default SettingsCharacteristic SpecificationDateaM
Factory Default SettingsPropaq Encore Reference Guide 159 Welch AllynDisplay Wave Select All waves are on except NIBPTrend Group NIBPAlarm Limits All
Introducing the Propaq EncoreWelch Allyn 16 Propaq Encore Reference GuideExpansion ModuleThe Propaq Encore Expansion Module attaches to the monitor an
In-Service Simulated ValuesWelch Allyn 160 Propaq Encore Reference GuideIn-Service Simulated ValuesIn-Service ValuesChannel Display Initial Value Alte
Propaq Encore Reference Guide 161 Welch AllynGlossary∆TDifference temperature. The difference between T1 and T2.AAMIAssociation for the Advancement of
Welch Allyn 162 Propaq Encore Reference GuideAuto IntervalThe interval at which NIBP measurements are initiated when operating in the automatic mode.B
Propaq Encore Reference Guide 163 Welch AllynCO2A patient channel indicating the by-product of respiration, carbon dioxide, which is exhaled by the lu
Welch Allyn 164 Propaq Encore Reference GuideEquipment AlertOccurs when the Propaq Encore detects an equipment condition requiring operator assistance
Propaq Encore Reference Guide 165 Welch AllynGas CompensationA correction factor required to obtain accurate CO2 readings when elevated levels of O2 o
Welch Allyn 166 Propaq Encore Reference GuideIn-service ModeA user training aid built into all Propaq Encores that provides simulated signals for all
Propaq Encore Reference Guide 167 Welch AllynPAPulmonary artery (label for an invasive blood pressure channel).ParameterSee Vital Sign Parameter.Patie
Welch Allyn 168 Propaq Encore Reference GuideRescale ModeThe method used in invasive pressure display to show each waveform against its own scale. The
Propaq Encore Reference Guide 169 Welch AllynSYNCSynchronization. Two uses apply:•A digital output pulse from the right side panel that starts within
Introducing the Propaq EncorePropaq Encore Reference Guide 17 Welch AllynImpedance Pneumography (RESP) OptionThe RESP option detects the rate or absen
Welch Allyn 170 Propaq Encore Reference GuideWindowAn area on the display screen in which information is displayed.YSIAn acronym for Yellow Springs In
Welch Allyn 171 Propaq Encore Reference GuideIndexAac adapter connector 18AccessoriesProducts & Accessories book 13Acuity System 115Acuity menu 11
IndexPropaq Encore Reference Guide 172 Welch AllynCO2 Menu 1mm/sec button 78, 82RANGE button 78, 81CO2 Menu 2FLOWRATE button 79GAS COMP button 78, 79,
IndexWelch Allyn 173 Propaq Encore Reference GuideESIS 40Expansion ModuleCO2 16Printer 16SpO2 16Extended Mode 40, 46Extended storage 127FFactory defau
IndexPropaq Encore Reference Guide 174 Welch AllynNIBP button 60, 61SETUP button 29SpO2/CO2 button 26MAIN MENU button 25Mainstream Capnography (see CO
IndexWelch Allyn 175 Propaq Encore Reference GuideAUTO/MAN button 61INTERVAL button 61START button 60STOP button 60TURBOCUF button 61NIBP TICKET setti
IndexPropaq Encore Reference Guide 176 Welch AllynPRINT button (see also Trends Menu) 113Printer Setup window 110RESP waveforms 108Setup Menu 2 109spe
IndexWelch Allyn 177 Propaq Encore Reference GuideSensor selection 69, 70size indicator 107specifications 144SpO2 Menus 69, 71Spot-Check monitoring 73S
IndexPropaq Encore Reference Guide 178 Welch AllynSpO2 70, 71Waveformsdisplay priorities 31ECG 31scales 54, 81size 45WindowsPropaq screen 24ZZERO (P1/
Using the Propaq EncoreWelch Allyn 18 Propaq Encore Reference GuideUsing the Propaq EncoreSystem Controls (Right Side Panel)On/Off SwitchThis switch t
Using the Propaq EncorePropaq Encore Reference Guide 19 Welch AllynDefib Sync ConnectorThis connector allows connection with a LIFEPAK 5 or LIFEPAK 6s
Copyright © 2004 by Welch Allyn. Welch Allyn ® is a registered trademark of Welch Allyn. Propaq ® , Propaq Encore ® , and Acuity ® are registered
Using the Propaq EncoreWelch Allyn 20 Propaq Encore Reference GuidePower-up Screen1. If the Propaq Encore has been used for a previous patient, switch
Using the Propaq EncorePropaq Encore Reference Guide 21 Welch AllynPowerup Equipment Alert: Program Fault, Settings LostIf a PROGRAM FAULT: SETTINGS L
Using the Propaq EncoreWelch Allyn 22 Propaq Encore Reference GuidePatient ConnectionsThe left side panels differ depending on the Propaq Encore model
Using the Propaq EncorePropaq Encore Reference Guide 23 Welch AllynOption ConnectorsMainstream CO2 ConnectorSidestream CO2 ConnectorNELLCORSENSORS/CAP
Using the Propaq EncoreWelch Allyn 24 Propaq Encore Reference GuidePropaq Encore DisplayThe display shows waveforms, vital sign numeric values, Propaq
Using the Propaq EncorePropaq Encore Reference Guide 25 Welch AllynYou can select up to three waveforms to be shown on the Propaq. When only one wavef
Using the Propaq EncoreWelch Allyn 26 Propaq Encore Reference GuideThe five buttons below the screen, and their associated labels located on the screen
Using the Propaq EncorePropaq Encore Reference Guide 27 Welch AllynKey-press Route to Setup Menu 1STAT SETSUSPENDALL ALRM LIMITSPREV MENUSETUPNIBP ECG
Using the Propaq EncoreWelch Allyn 28 Propaq Encore Reference GuideKey-press Route to Setup Menu 2 ON/OFFNEXTINSERVPREV MENUMORENIBP TEST IBP TEST SpO
Monitor SetupPropaq Encore Reference Guide 29 Welch AllynMonitor SetupSetup Menu 1 is accessed by pressing the SETUP button on the Main Menu. STATSCAL
Welch Allyn 3 Propaq Encore Reference Guide Contents General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Monitor SetupWelch Allyn 30 Propaq Encore Reference GuideHR/PR TONE Sets the Heart Tone volume to HIGH, MEDIUM, LOW, or OFF.PATIENT MODE Pressing CHAN
Monitor SetupPropaq Encore Reference Guide 31 Welch AllynSelecting Waveforms for DisplayTo select waveforms for display, press SETUP, WAVE SEL. Use th
Monitor SetupWelch Allyn 32 Propaq Encore Reference GuideTime/Day Settings and TrendsWhen you change the hour/minute/second setting for the monitor in
Monitor SetupPropaq Encore Reference Guide 33 Welch AllynSetting the Current, Custom, and Powerup ModesThe Propaq Encore has two sets of patient mode
Monitor SetupWelch Allyn 34 Propaq Encore Reference GuideFrom the Main Menu, press SETUP, MORE, CHANGE, SETUP. The Mode Setup window appears:The aster
Printer FunctionsPropaq Encore Reference Guide 35 Welch AllynPrinter FunctionsPress SETUP, MORE, PRINTER to display the printer menu and setup window.
Printer FunctionsWelch Allyn 36 Propaq Encore Reference GuidePRINTER FAULT MessagesThese PRINTER FAULT messages can appear in an equipment alert windo
Learning the Propaq EncorePropaq Encore Reference Guide 37 Welch AllynLearning the Propaq EncoreUsing In-Service ModeYou can practice using the Propaq
Learning the Propaq EncoreWelch Allyn 38 Propaq Encore Reference GuideNIBPFor noninvasive pressure measurements, keep the Propaq Encore in manual NIBP
Propaq Encore Reference Guide 39 Welch AllynPatient MonitoringECG/RESPIntended Use–Impedance Pneumography (RESP) The Respiration channel is intended t
ContentsPropaq Encore Reference Guide 4 Welch Allyn What You Can Do With In-Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ECG/RESPWelch Allyn 40 Propaq Encore Reference GuideIntended Use–ECGThe Propaq Encore is intended for ECG monitoring of either a five-lead or three-lea
ECG/RESPPropaq Encore Reference Guide 41 Welch AllynEven though the Propaq Encore contains fully isolated patient-connected circuitry, it has not been
ECG/RESPWelch Allyn 42 Propaq Encore Reference GuidePreparationPreparing for ECG monitoring with the Propaq Encore requires you to prepare the monitor
ECG/RESPPropaq Encore Reference Guide 43 Welch Allyn4. Apply the electrodes to the patient.5. Support the ECG cable so it does not stress the electrod
ECG/RESPWelch Allyn 44 Propaq Encore Reference Guide6. If an electrosurgical unit is going to be used, place the ECG cable and electrode wires as far
ECG/RESPPropaq Encore Reference Guide 45 Welch AllynHow ECG/RESP is DisplayedBecause of the critical nature of monitoring ECG, it is always displayed
ECG/RESPWelch Allyn 46 Propaq Encore Reference GuideWhen you press MORE in the first ECG/RESP Menu, a status window appears showing you the current ECG
ECG/RESPPropaq Encore Reference Guide 47 Welch AllynUsing the Propaq Encore With Pacemaker PatientsPacemaker SignalsIf the patient being monitored has
ECG/RESPWelch Allyn 48 Propaq Encore Reference GuideNoise on the signalNoise on the ECG signal may be detected as pacer signals, causing the pacer ind
ECG/RESPPropaq Encore Reference Guide 49 Welch AllynRESP Messages LEAD FAIL. One or more electrodes are making very poor or no contact. Check for prop
ContentsWelch Allyn 5 Propaq Encore Reference Guide CO2 Measurements and Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Invasive PressureWelch Allyn 50 Propaq Encore Reference GuideInvasive PressureThis section applies only to Propaq Encore Models 204EL and 206EL. If yo
Invasive PressurePropaq Encore Reference Guide 51 Welch AllynPreparationPreparing for invasive pressure monitoring with the Propaq Encore requires you
Invasive PressureWelch Allyn 52 Propaq Encore Reference Guide3. Wait for a tone to briefly sound and the word ZEROED to appear in the blood pressure nu
Invasive PressurePropaq Encore Reference Guide 53 Welch AllynTo remove the ZERO REJECTED message and to restore the invasive pressure numerics during
Invasive PressureWelch Allyn 54 Propaq Encore Reference GuideInvasive Pressure MessagesThe following messages can appear in the numerics window.NOT ZE
Invasive PressurePropaq Encore Reference Guide 55 Welch AllynZERO REJECTED (or REJECT). Unable to establish a zero reference value. The message remain
NIBPWelch Allyn 56 Propaq Encore Reference GuideNIBPIntended UseThe Propaq Encore noninvasive blood pressure channel (NIBP) indirectly measures arteri
NIBPPropaq Encore Reference Guide 57 Welch AllynNIBP measurements are affected by normal physiological pressure variations from reading to reading. No
NIBPWelch Allyn 58 Propaq Encore Reference GuideIf artifact is so severe while Smartcuf is enabled that it affects the accuracy of an NIBP measurement
NIBPPropaq Encore Reference Guide 59 Welch AllynFor information about patient mode specifications, see “NIBP Specifications” on page 142.PreparationSett
ContentsPropaq Encore Reference Guide 6 Welch Allyn Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
NIBPWelch Allyn 60 Propaq Encore Reference Guide3. Screw the hose connector onto the NIBP connector on the monitor's left side. 4. If motion arti
NIBPPropaq Encore Reference Guide 61 Welch AllynNIBP Displayed in Waveform Window AUTO/MANThis button switches the mode between Automatic or Manual Mo
NIBPWelch Allyn 62 Propaq Encore Reference GuideSet the NIBP Alarm LimitsSet the alarm limits according to your hospital's standards.NIBP Display
NIBPPropaq Encore Reference Guide 63 Welch AllynAIR LEAK, CHECK HOSE (ERR# 1). The Propaq Encore could not properly inflate cuff. Check the hose and cu
NIBPWelch Allyn 64 Propaq Encore Reference GuideLOW BATTERY, NIBP DISABLED (ERR# 11). The battery lacks sufficient voltage to be able to operate the NI
NIBPPropaq Encore Reference Guide 65 Welch AllynNIBP IN PROGRESS MessageThe monitor displays this message when noise or artifact such as vehicle motio
TemperatureWelch Allyn 66 Propaq Encore Reference GuideTemperatureIntended UsePropaq Encore monitors provide two temperature channels (except for the
TemperaturePropaq Encore Reference Guide 67 Welch AllynTemperature MessagesThe following messages can appear in an equipment alert window. A temperatu
Pulse Oximetry (SpO2)Welch Allyn 68 Propaq Encore Reference GuidePulse Oximetry (SpO2) Each SpO2 sensor is designed for application to a specific site
Pulse Oximetry (SpO2)Propaq Encore Reference Guide 69 Welch AllynIf NIBP will be monitored while using SpO2, place the NIBP cuff on a different limb t
ContentsWelch Allyn 7 Propaq Encore Reference Guide Extended Storage Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pulse Oximetry (SpO2)Welch Allyn 70 Propaq Encore Reference Guide4. Press SIZE to adjust the waveform size for best viewing (1x, 2x, 4x, or 8x).5. Adj
Pulse Oximetry (SpO2)Propaq Encore Reference Guide 71 Welch Allyn4. From the Main Menu, press SpO2 (or SpO2/CO2, then SpO2) to display the first SpO2 m
Pulse Oximetry (SpO2)Welch Allyn 72 Propaq Encore Reference Guide9. If the C-LOCK function is desired, press C-LOCK to set it to ON.10. Set alarm limi
Pulse Oximetry (SpO2)Propaq Encore Reference Guide 73 Welch AllynPerform SpO2 “Spot-Check” MonitoringThe SpO2 Standby Mode allows you to remove the Sp
Pulse Oximetry (SpO2)Welch Allyn 74 Propaq Encore Reference GuideSTANDBY is displayed in the numeric window when the SpO2 sensor is disconnected from
Capnography (CO2)Propaq Encore Reference Guide 75 Welch AllynCapnography (CO2)Intended UseThe Propaq’s Capnography (CO2) option is intended to noninva
Capnography (CO2)Welch Allyn 76 Propaq Encore Reference GuideMainstream CO2 OptionThe Mainstream CO2 option measures the carbon dioxide content of a p
Capnography (CO2)Propaq Encore Reference Guide 77 Welch AllynDisplayed values of ETCO2 and INCO2 are the highest and lowest values (respectively) of C
Capnography (CO2)Welch Allyn 78 Propaq Encore Reference GuideUNCAL indicates the monitor has detected a problem such as a lack of calibration, an obst
Capnography (CO2)Propaq Encore Reference Guide 79 Welch AllynSidestream CO2 Menu and Status WindowWhen Sidestream CO2 is active, press MORE from the
ContentsPropaq Encore Reference Guide 8 Welch Allyn Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Capnography (CO2)Welch Allyn 80 Propaq Encore Reference GuideMainstream CO2 Monitoring1. Select the appropriate airway adapter.Use only accessories pr
Capnography (CO2)Propaq Encore Reference Guide 81 Welch Allyn3. Plug in the CO2 sensor cable to the Mainstream CO2 connector on the Propaq left side p
Capnography (CO2)Welch Allyn 82 Propaq Encore Reference Guide3. Press mm/s to select either 3.13, 6.25 or 12.5 mm/sec. The default setting is 6.25 mm/
Capnography (CO2)Propaq Encore Reference Guide 83 Welch AllynMotion artifact or other factors can cause false RR/BR alarms. To help minimize false ala
Capnography (CO2)Welch Allyn 84 Propaq Encore Reference Guide3. Connect to a non-intubated patient.OR4. Connect to an intubated patient.Use only acces
Capnography (CO2)Propaq Encore Reference Guide 85 Welch AllynConnect to a Non-Intubated Patient1. Position the cannula on the patient according to the
Capnography (CO2)Welch Allyn 86 Propaq Encore Reference GuideConnect to an Intubated Patient1. Connect the gas sampling elbow and elbow connector into
Capnography (CO2)Propaq Encore Reference Guide 87 Welch AllynALTIMETER FAILURE - RATE. The altimeter has detected that the ambient pressure is changin
Capnography (CO2)Welch Allyn 88 Propaq Encore Reference GuideSidestream CO2 MessagesEquipment messages for the Sidestream CO2 option can appear on the
Capnography (CO2)Propaq Encore Reference Guide 89 Welch AllynOCCLUSION - CHECK WATERTRAP/TUBING. Blockage has been detected on the Sidestream CO2 inpu
Propaq Encore Reference Guide 9 Welch Allyn General Information Safety Summary This Safety Summary should be read by all Propaq Encore users. Specific
Capnography (CO2)Welch Allyn 90 Propaq Encore Reference Guide
Propaq Encore Reference Guide 91 Welch AllynAlarms and LimitsAlarms Window and MenuYou can set the alarms in normal operating mode or in-service mode.
Alarms Window and MenuWelch Allyn 92 Propaq Encore Reference GuideTo adjust the alarm tone volume, press SETUP, then MORE. The status window appears.
Types of Propaq AlarmsPropaq Encore Reference Guide 93 Welch AllynTypes of Propaq AlarmsPatient Alarms: Definitions and IndicationsThe Propaq Encore ca
Types of Propaq AlarmsWelch Allyn 94 Propaq Encore Reference GuideThe alarm tones are summarized in the table:Life-Threatening AlarmsLife-threatening
Types of Propaq AlarmsPropaq Encore Reference Guide 95 Welch AllynResponding to Patient AlarmsSuspend an Alarm ToneYou can temporarily turn off an ala
Types of Propaq AlarmsWelch Allyn 96 Propaq Encore Reference Guide• Manually change the violated alarm limit value by selecting the violated limit val
Alarm HoldoffsPropaq Encore Reference Guide 97 Welch AllynAlarm HoldoffsPropaq Audio and Visual Alarm HoldoffsMotion artifact or other factors can cau
Setting Alarm LimitsWelch Allyn 98 Propaq Encore Reference GuideSetting Alarm LimitsAdjust Limits with STAT SETWhen you want to quickly set all alarm
Setting Alarm LimitsPropaq Encore Reference Guide 99 Welch AllynTurning On and Off All LimitsIf you want to turn all limits on or off without changing