Packet Structure SureTemp Plus Module OEM Implementation
34 Confidential Welch Allyn OEM Technologies
Complex Data Types (Structures)
TEMP_STATUS
Type Description Values/Units
uint16 POST status 00: No errors
255: POST error code not yet ready
see Error Table (page 71) for additional
codes
uint16 POST Tests The bit mapping for the tests.
0: not performed, 1: performed
15: Data Bus Test
14: Address Bus Test
13: Device Physical Memory Address Test
12: Segments Test
11: Thermistor Test
10: Heater Test
09: Spare
08: Spare
07: Spare
06: Spare
05: Spare
04: Spare
03: Spare
02: Spare
01: Spare
00: Spare