Commands and Responses SureTemp Plus Module OEM Implementation
46 Confidential Welch Allyn OEM Technologies
(3109) Erase Event Log
Causes the module to erase the event log and return it to the empty state.
Command Code (hex)
3109
Data
Size (bytes) Type Description
0 N/A None
Response Data
Size (bytes) Type Description
0 N/A None
(3503) Lock Thermometer Probe
Locks or unlocks access to the module’s Thermometer probe switch. Once locked, the probe
switch can be turned on or off via the “Set Temperature Probe” command.
Command Code (hex)
3503
Data
Size (bytes) Type Description
1 boolean 0: Unlocked
1: Locked
Response Data
Size (bytes) Type Description
0 N/A None