
Commands and Responses SureTemp Plus Module OEM Implementation
42 Confidential Welch Allyn OEM Technologies
(2A12) Set Temperature Method
Causes the module to set the current temperature method from the host, forcing the module
into the host-specified mode.
Command Code (hex)
2A12
Data
Size (bytes) Type Description
1 int8 0: Normal (Predictive)
1: Monitor
Response Data
Size (bytes) Type Description
0 N/A None
Notes
This command returns an “Invalid operation in this context” error if the temperature probe is
out of the probe well, and returns a “Data Range” error if the specified temperature method is
undefined.
This command has effect only in a predict cycle, and can be used only to transition the
monitor from predict mode to monitor mode.
Commentaires sur ces manuels