96 Repair procedures Welch Allyn Propaq CS Vital Signs Monitor
Installing the printer
1. Lay the Chassis on its back
2. Completely open the paper door.
3. Slide the Printer Assembly into the guides as shown below, making sure that after the
door latch mechanism clears the chassis you rotate it clockwise to engage the printer
door pin. (Also see the figure above.)
printer assy in stallation
4. Carefully close and open the door a couple times to check that the door pin and latch
mechanism function without binding.
5. Connect the battery cable and printer cable to the Printer Main Board. It is important
to properly dress the cables as shown below.
‘printer cable dress
6. Connect the button cable to the front of the Printer Main Board.
7. Install the Expansion Module front panel, close the Expansion Module, and reattach it
to the monitor.
Caution As you install the Printer Assembly, rotate the printer door latch
mechansim out of the way so it clears the Chassis. The Printer Assembly fits
snugly, but should not require much force to slide it in place.
Guides
0696-51
Battery
Cable
(Dress tight
against plastic)
Battery
Cable
Printer
Cable