Welch Allyn PROPAQ CS Manuel de service Page 96

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 142
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 95
92 Repair procedures Welch Allyn Propaq CS Vital Signs Monitor
Closing Expansion Module
1. Make sure all cables are in place and the gaskets removed earlier are available.
2. Set the rubber gaskets in their proper places as shown in the figure below.
3. Connect the printer cable and expansion cable to the SCP or MSP Board.
expansion rear panel
4. Set the Chassis onto the rear panel, confirming that the gaskets don’t slip out of a
channel and become pinched between the rear panel and the Chassis.
5. Torque screws to 4.75 to 5.25 in-lb (0.54 to 0.59 N-m).
Caution To prevent damage to the Expansion Module Chassis, make sure that
the spacers are in place. The Chassis may crack if screws are tightened without
the spacers in place. Note that the five upper spacers are not used when the
SpO
2
module is present.
0696-47
Gasket
Material
Vue de la page 95
1 2 ... 91 92 93 94 95 96 97 98 99 100 101 ... 141 142

Commentaires sur ces manuels

Pas de commentaire