138 Fluke patient simulator modification Welch Allyn Propaq CS Vital Signs Monitor
Incompatibilities with other simulators may also cause erroneous readings and/or missing
ECG. Check Propaq CS monitor readings with the simulator against a known source
before using the simulator for calibration.
Function Color Pin No. Propaq CS 244/246
Output (+) Black 4 C
Output (-) Red 1 A & E
Exciter (+) White 3 B
Exciter (-) Green 5 D
ECG Ref Blue 2 No Connection
810-0696-57
W1
W2